Research on Key Projects of Electromagnetic Compatibility Field Testing for Large Medical Robot

Journal: Journal of Clinical Medicine Research DOI: 10.32629/jcmr.v5i3.2634

Mingjian Zhu, Chengke Zhu, Qingyu Li, Shu Yang, Shuyi Zhang, Mei Wang, Deyu Li, Meili Liu, Meng Liu

Shandong Institute of Medical Device and Pharmaceutical Packaging Inspection, Jinan 250101, Shandong, China NMPA Key Laboratory for Quality Evaluation of Medical Materials and Biological Protective Devices, Jinan 250101, Shandong, China

Abstract

The present study aims to investigate the electromagnetic compatibility testing method for large medical robots. This paper explores the operational mode and fundamental performance of large medical robots in electromagnetic compatibility testing, based on the working principle and structural composition of a sampling robot. Furthermore, it focuses on analyzing the field detection methods for radiation emission test and radiation immunity test of large medical robots. Ultimately, this research provides valuable insights for conducting electromagnetic compatibility field tests on large medical robots.

Keywords

large medical robot, electromagnetic compatibility, field test, radiated emission test , radiated RF electromagnetic field immunity test

References

[1]QIU Qiu-fei,ZHOU Wu-yuan,WU Qiao-ling, et al. Research on Technical Development of Intelligent Medical Anti Epidemic Robot Based on Patent Information[J]. Machinery Design & Manufacture, 2023(10): 248-253, 257.
[2]XIE Neng,CAO Qi-xin,WANG Jin-wu, et al. Research on Control Points of Medical Robot[J]. China Medical Device Information, 2022, 28(12): 1-4, 66.
[3]JIANG Wei-ping,MENG Zi-han,JI Zhi-kang, et al. Design of robot aided visual angle detection system for medical screen based on visual servo[J]. China Measurement & Test, 2022, 48(05): 128-133, 141.
[4]LIU Mei-li,ZHU Ming-jian,WANG Xiao-hui, et al. Site Testing Research of Electromagnetic Compatibility for Large Steam Sterilizers[J]. China Medical Devices, 2020, 35(03): 21-24.
[5]WANG Quan,SU Zong-wen,LI Shu, et al. Field Electromagnetic Compatibility Testing for Large Medical Electrical Equipment[J]. China Medical Devices, 2015, 30(09): 70-71, 11.
[6]HUANG Dan,XU Ke. Field Test of Radiation Emission for Large Medical Molecular Sieve Oxygen Generating Equipment[J]. China Medical Device Information, 2018, 24(17): 12-13.
[7]Chen Ting, Yu Hua-tong, Zou Ming-ming, et al. Study of EMC On-site Test for Large Medical Equipment or System[J]. Safety & EMC, 2016(02): 25-27, 59.
[8]FENG Xin. Research on the Environment Elimination in EMC Site Test[D]. Chinese Master's Theses Full-text Database, 2013.
[9]Liu Yuan, CAI Wenjiang. Research on On-site Testing of Electromagnetic Radiation Disturbance[J]. Safety & EMC, 2010(05): 29-32, 68.
[10]Zhang Zhi-qing. Attention Issue in Test of Large-Scale Medical Electrical Equipment[J]. China Medical Device Information, 2009, 15(06): 43-45.

Copyright © 2024 Mingjian Zhu, Chengke Zhu, Qingyu Li, Shu Yang, Shuyi Zhang, Mei Wang, Deyu Li, Meili Liu, Meng Liu

Creative Commons License
This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License